User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 2521 Verification Test (Ver1.4) DATE: 03-Oct-93 AUTHOR: User Contributed REVISION: 1.4 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 50 NUMBER OF LINES: 98 CONFIGURATION: Fluke 5700A ============================================================================= * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Source: Service Manual PM P/N: @ * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * TUR Configuration: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * When the T.U.R. is < 2:1, the source instrument, e.g. 5700A, must be * characterized and appropriate corrections entered in the correction file * (e.g. 5700.COR). Where T.U.R.'s are > 2:1 but < 4:1, the test tolerance * is reduced by root sum square (rss) techniques. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * This procedure is not executable in MET/CAL versions below ver. 2.00 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Fluke make no warranty, expressed or implied, as to the fitness or * useability of this procedure. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Main Procedure TESTS: Verification * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * The following subprocedures are called from this main procedure: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DRAW 7 78 7 78 75 5 50 55 1.002 DISP Turn POWER on. [DRAW10,5,F] 1.003 DISP Allow the UUT to warm up for 30 min. and proceed the 1.003 DISP test at an ambient temp. of 23 +/- 1 deg. Celsius. 1.004 DISP Set the UUT to AUTO ranging. [DRAW10,70,F] 1.005 DISP Turn selector switch to V DC. [DRAW50,80,F] 1.006 ASK- R 1.007 DISP Shortcircuit the 0 and the V-Ohm-mA terminals to check 1.007 DISP the zero point (000.00 mV +/- 1 digit). 1.007 DISP Readjust the zero point if necessary (screwdriver- 1.007 DISP adjustment). [DRAW20,55,F] 1.008 DISP Remove the shortcircuit, and connect testleads to the 1.008 DISP 0 and the V-Ohm-mA terminals. 1.009 HEAD {DC VOLTAGE TEST} 1.010 5700 200 190mV 0.01/ 0.03% 2W 2.001 5700 200 -190mV 0.01/ 0.03% 2W 3.001 5700 2 1.9V 0.01/ 0.03% 2W 4.001 5700 2 -1.9V 0.01/ 0.03% 2W 5.001 5700 20 19V 0.01/ 0.03% 2W 6.001 5700 20 -19V 0.01/ 0.03% 2W 7.001 5700 200 190V 0.01/ 0.03% 2W 8.001 5700 200 -190V 0.01/ 0.03% 2W 9.001 5700 2000 1000V 0.01/ 0.03% 2W 10.001 5700 2000 -1000V 0.01/ 0.03% 2W 11.001 HEAD {AC VOLTAGE TEST} 11.002 DISP Turn selector switch to V AC. [DRAW50,80,F] 11.003 5700 200 19mV 0.15/ 0.3% 60H 2W 12.001 5700 200 190mV 0.15/ 0.3% 60H 2W 13.001 5700 200 190mV 0.15/ 1% 10kH 2W 14.001 5700 200 190mV 1/ 5% 100kH 2W 15.001 5700 2 1.9V 0.15/ 0.3% 60H 2W 16.001 5700 2 1.9V 0.15/ 1% 10kH 2W 17.001 5700 2 1.9V 1/ 5% 100kH 2W 18.001 5700 20 19V 0.15/ 0.3% 60H 2W 19.001 5700 20 19V 0.15/ 1% 10kH 2W 20.001 5700 20 19V 1/ 5% 100kH 2W 21.001 5700 200 190V 0.15/ 0.3% 60H 2W 22.001 5700 200 190V 0.15/ 0.3% 10kH 2W 23.001 5700 200 190V 1/ 5% 100kH 2W 24.001 5700 2000 600V 0.15/ 0.3% 50H 2W 25.001 HEAD {DIRECT CURRENT TEST} 25.002 DISP Turn selector switch to A DC. [DRAW50,80,F] 25.003 5700 20 19uA 0.05/ 0.2% 2W 26.001 5700 20 -19uA 0.05/ 0.2% 2W 27.001 5700 200 190uA 0.05/ 0.2% 2W 28.001 5700 200 -190uA 0.05/ 0.2% 2W 29.001 5700 2 1.9mA 0.05/ 0.2% 2W 30.001 5700 2 -1.9mA 0.05/ 0.2% 2W 31.001 5700 20 19mA 0.05/ 0.2% 2W 32.001 5700 20 -19mA 0.05/ 0.2% 2W 33.001 DISP Disconnect testlead from V-Ohm-mA terminal and connect 33.001 DISP it to the 0.2A-10A terminal. 33.002 5700 200 190mA 0.05/ 0.2% 2W 34.001 5700 200 -190mA 0.05/ 0.2% 2W 35.001 5700 2 1.9A 0.05/ 0.2% 2W 36.001 5700 2 -1.9A 0.05/ 0.2% 2W 37.001 HEAD {ALTERNATING CURRENT TEST} 37.002 DISP Turn selector switch to A AC. [DRAW50,80,F] 37.003 5700 2 1.9A 0.15/ 0.4% 100H 2W 38.001 5700 200 190mA 0.15/ 0.4% 100H 2W 39.001 DISP Disconnect testlead from 0.2A-10A terminal and connect 39.001 DISP it to the V-Ohm-mA terminal. 39.002 5700 20 19mA 0.15/ 0.4% 100H 2W 40.001 5700 2 1.9mA 0.15/ 0.4% 100H 2W 41.001 5700 200 190uA 0.15/ 0.4% 100H 2W 42.001 5700 20 19uA 0.15/ 0.4% 100H 2W 43.001 HEAD {RESISTANCE TEST} 43.002 DISP Turn selector switch to Ohms. [DRAW50,80,F] 43.003 5700 200 190Z 0.1/ 0.2% 2W 44.001 5700 2 1.9kZ 0.1/ 0.2% 2W 45.001 5700 20 19kZ 0.1/ 0.2% 2W 46.001 5700 200 190kZ 0.1/ 0.2% 2W 47.001 5700 2 1.9MZ 0.1/ 1% 2W 48.001 5700 20 19MZ 0.1/ 1% 2W 49.001 HEAD {DIODE TEST} 49.002 ASK- W 49.003 DISP Turn selector switch to DIODE. 49.004 5700 2 1.9kZ 0.1/ 0.5% 2W 50.001 DISP { } 50.001 DISP { } 50.001 DISP { } 50.001 DISP { } 50.001 DISP { } 50.001 DISP { } 50.001 DISP { } 50.002 END